Skip to content

MEP protocol

Running an assembled MEP. To build one, follow the MEP guide. Source documents: Calibration and Operation.

Every mode uses the same hardware: stirring on PWM 1, waste (product) on PWM 2, media on PWM 3, the CO₂ relay on PWM 4, and electrolysis on LED D at 2.5% from the electroPioreactor job.

Do these once per unit before the first experiment, and check the electrolysis values again before and after each experiment. A significant change means recalibrating.

  1. Run Pioreactor’s self test.
  2. Calibrate stirring.
  3. Calibrate the pumps and set the level: done in the guide’s Calibrate the pumps and set the level step.
  4. Make an OD600 standard curve on the Protocols page (device od90), from vials of known OD600 measured on a benchtop spectrophotometer plus a media-only blank. Redo it every 6 months, or whenever the optics change.
  5. Calibrate the CO₂ flow: done in the guide’s Calibrate CO₂ flow step.
  6. Record the electrolysis voltage and current at 2.5%: done in the guide’s Test electrolysis step.
Mode Media in Waste out Use it to
Batch No No Grow a culture on one fill until something runs out. CO₂, H₂ and O₂ are still fed, so strictly it is fed-batch.
Chemostat Fixed rate Fixed rate Hold a steady state at a constant dilution rate.
Turbidostat When OD is above target When OD is above target Hold a steady state at a target OD.
  1. Check the electrodes are at the standard depth, red on the anode and black on the cathode.
  2. Fill the vial with 14 ml of nutrient solution (MC02) and screw the cap down.
  3. Seat the vial in the Pioreactor.
  4. In the web interface, create an experiment (for example ed04_batch_<date>) and assign the unit to it.
  5. Run OD reading until the trace is stable, and save the OD blank for this experiment. OD reading will not start with OD calibrations enabled until a blank exists.
  6. Inoculate through one of the gas outlet luers with a syringe. The vial stays seated.
  7. Start, in this order: stirring (check the stir bar turns), OD reading, then electroPioreactor (electrolysis at 2.5%, sparging 3 seconds every hour). Bubbles should appear within 30 seconds.
  8. Watch OD rise. Record the electrolysis voltage at least daily, ideally hourly; the cathode should keep bubbling faster than the anode.
  9. At the endpoint, stop electroPioreactor and stirring, and sample through a gas outlet luer.
  10. Drain the vial, rinse it with distilled water, and clean it.

Start here only after a successful batch run.

  1. Set up as for batch, up to inoculation.
  2. Set the waste line to your level (see the guide’s Calibrate the pumps and set the level step). The waste pump over-runs each dilution by design, so the line’s height is the level.
  3. In the dosing automation, choose chemostat, with an exchange volume (for example 0.5 ml) and a duration between doses. The dilution rate D (h⁻¹) = (exchange volume × 60 / duration) / 14.
  4. Start stirring, OD reading, electroPioreactor and the dosing automation.
  5. Over many hours OD settles to a steady state if D is below the culture’s maximum growth rate; otherwise the culture washes out.
  6. End and strip down as for batch.
  1. Set up as for chemostat.
  2. In the dosing automation, choose turbidostat: target biomass as expected for your culture (for example 0.5), biomass signal left at auto, and exchange volume 1 to 2 ml.
  3. Start stirring, OD reading, electroPioreactor and the dosing automation.
  4. OD becomes a sawtooth around the target as the unit doses whenever OD crosses it.
  5. End and strip down as for batch.

In the experiment setup: name, mode, unit, dates, media batch, inoculum source and volume. As you go: electrolysis voltage and current at the start, at least daily and at the end, and anything unusual (a loose electrode bolt, a change in bubble rate, anode discolouration, level drift, a knocked unit).