Set up the CO₂ flow test
Your setup: — change
Set up and check the test here; Calibrate CO₂ flow, next, measures and adjusts the flow. A wrong Drops the cylinder's pressure (about 60 bar when full) to a working pressure, about 1 bar here. Its flathead screw fully anti-clockwise is closed. pressure found after that would mean measuring again.
- Set the regulator to 1 bar.
- Cap one of the two gas outlets with a A standard medical screw fitting. A male luer (the spigot) screws into a female luer (the socket). In this build, male ends mark outlets. cap.
- Run 1/16” tubing from the open gas outlet into a water bath, under a measuring cylinder full of water.
- Start the electroPioreactor job, or turn on the relay in the Pioreactor UI.
- Bubbling gas through the culture. Here, CO₂ from the cylinder flows down the anode on a timer set in the electroPioreactor job. until the water’s CO₂ concentration has equilibrated (assumed, or measured if you can).
Parts
- 1×CO₂ regulatorearlier step
- 1×Luer lock capearlier step
- 1×Silicone tubing, 50 cmearlier step
Check your work
What does the Drops the cylinder's pressure (about 60 bar when full) to a working pressure, about 1 bar here. Its flathead screw fully anti-clockwise is closed.'s outlet gauge read?
Screw the regulator in until it reads 1 bar, before touching the A fine-adjustment valve on the solenoid's front port. It sets how fast CO₂ flows; turn clockwise to close..
Turn the regulator anti-clockwise until it reads 1 bar, before touching the needle valve.
Show the answer
Answer: About 1 bar
- If you see Well under 1 bar: Screw the regulator in until it reads 1 bar, before touching the A fine-adjustment valve on the solenoid's front port. It sets how fast CO₂ flows; turn clockwise to close..
- If you see Well over 1 bar: Turn the regulator anti-clockwise until it reads 1 bar, before touching the needle valve.
While Bubbling gas through the culture. Here, CO₂ from the cylinder flows down the anode on a timer set in the electroPioreactor job., where does the gas leave the vial?
Cap one gas outlet with a A standard medical screw fitting. A male luer (the spigot) screws into a female luer (the socket). In this build, male ends mark outlets. cap, or the measuring cylinder catches only part of the flow.
Check the job is sparging (or the relay is on) and the A fine-adjustment valve on the solenoid's front port. It sets how fast CO₂ flows; turn clockwise to close. is open.
Show the answer
Answer: Only through the open outlet, into the measuring cylinder
- If you see Through both outlets: neither is capped: Cap one gas outlet with a A standard medical screw fitting. A male luer (the spigot) screws into a female luer (the socket). In this build, male ends mark outlets. cap, or the measuring cylinder catches only part of the flow.
- If you see Nowhere: nothing leaves the open outlet: Check the job is sparging (or the relay is on) and the A fine-adjustment valve on the solenoid's front port. It sets how fast CO₂ flows; turn clockwise to close. is open.
Step checked