Set up the CO₂ flow test

Set up and check the test here; Calibrate CO₂ flow, next, measures and adjusts the flow. A wrong Drops the cylinder's pressure (about 60 bar when full) to a working pressure, about 1 bar here. Its flathead screw fully anti-clockwise is closed. pressure found after that would mean measuring again.

  1. Set the regulator to 1 bar.
  2. Cap one of the two gas outlets with a A standard medical screw fitting. A male luer (the spigot) screws into a female luer (the socket). In this build, male ends mark outlets. cap.
  3. Run 1/16” tubing from the open gas outlet into a water bath, under a measuring cylinder full of water.
  4. Start the electroPioreactor job, or turn on the relay in the Pioreactor UI.
  5. Bubbling gas through the culture. Here, CO₂ from the cylinder flows down the anode on a timer set in the electroPioreactor job. until the water’s CO₂ concentration has equilibrated (assumed, or measured if you can).

Parts

  • 1×CO₂ regulatorearlier step
  • 1×Luer lock capearlier step
  • 1×Silicone tubing, 50 cmearlier step

Check your work

  1. What does the Drops the cylinder's pressure (about 60 bar when full) to a working pressure, about 1 bar here. Its flathead screw fully anti-clockwise is closed.'s outlet gauge read?

    Show the answer

    Answer: About 1 bar

    • If you see Well under 1 bar: Screw the regulator in until it reads 1 bar, before touching the A fine-adjustment valve on the solenoid's front port. It sets how fast CO₂ flows; turn clockwise to close..
    • If you see Well over 1 bar: Turn the regulator anti-clockwise until it reads 1 bar, before touching the needle valve.
  2. While Bubbling gas through the culture. Here, CO₂ from the cylinder flows down the anode on a timer set in the electroPioreactor job., where does the gas leave the vial?

    Show the answer

    Answer: Only through the open outlet, into the measuring cylinder

    • If you see Through both outlets: neither is capped: Cap one gas outlet with a A standard medical screw fitting. A male luer (the spigot) screws into a female luer (the socket). In this build, male ends mark outlets. cap, or the measuring cylinder catches only part of the flow.
    • If you see Nowhere: nothing leaves the open outlet: Check the job is sparging (or the relay is on) and the A fine-adjustment valve on the solenoid's front port. It sets how fast CO₂ flows; turn clockwise to close. is open.