Calibrate CO₂ flow

The Drops the cylinder's pressure (about 60 bar when full) to a working pressure, about 1 bar here. Its flathead screw fully anti-clockwise is closed. was checked at 1 bar in Set up the CO₂ flow test.

  1. Record how long the CO₂ takes to fill the measuring cylinder.
  2. Calculate the flow rate.
  3. Adjust the A fine-adjustment valve on the solenoid's front port. It sets how fast CO₂ flows; turn clockwise to close., and repeat until you reach the target flow rate.
Shutting the gas off after a session
  1. Turn the regulator anti-clockwise until its screw hangs loose: it is then fully off.
  2. Close the adapter’s pin.
  3. Turn the The dial on the front of the solenoid. 0 is normal (closed without power); 1 holds the valve open. to 1 to vent the line to zero, then back to 0.

Parts

  • 1×CO₂ needle valveearlier step

Check your work

  1. How does the measured flow rate compare with your target?

    Show the answer

    Answer: On target, within your tolerance

    • If you see Above target, beyond your tolerance: Close the A fine-adjustment valve on the solenoid's front port. It sets how fast CO₂ flows; turn clockwise to close. a little (clockwise), and repeat the measurement until it matches.
    • If you see Below target, beyond your tolerance: Open the needle valve a little (anti-clockwise), and repeat the measurement until it matches.