Calibrate the pumps and set the level
The pump channels and the 12 V supply were checked in Nutrient solution.
- Fill the media bottle with water.
- Calibrate the pumps on the Protocols page, waste on One of the HAT's five switched power outputs. In this build: 1 stirring, 2 media pump, 3 waste (product) pump, 4 CO₂ solenoid (relay), 5 heating. 2 and media on PWM 3: catch each dispense in a tared weighing boat on the analytical balance and enter what it weighed (1 g of water is about 1 ml).
- Unscrew the cap, empty, rinse and dry the vial, and fill it to 14 ml with water.
- Screw the cap back on, and weigh the vial to confirm the volume.
- Adjust the Media Out tube so it is only just submerged at that level.
- Run the media and waste pumps, then weigh the vial again.
- Repeat from item 5 until the vial holds 14 ml after running the pumps.
Why the Media Out tube sets the level
The waste pump over-runs each dilution by design, so the height of the Media Out tube is the level setpoint, not the ratio of media to waste volume. See forum #801.
Parts
- 2×12 V peristaltic pumpsearlier step
- 2×250 ml GL45 borosilicate bottleearlier step
- 1×Pioreactor 20 ml glass vialearlier step
- 1×O-ring vial capearlier step
Tools
- 1×Analytical balancetool
Check your work
Which pumps have you calibrated?
Calibrate the media pump (One of the HAT's five switched power outputs. In this build: 1 stirring, 2 media pump, 3 waste (product) pump, 4 CO₂ solenoid (relay), 5 heating. 3) too.
Calibrate the waste pump (PWM 2) too.
Show the answer
Answer: The waste pump and the media pump
- If you see The waste pump, not the media pump: Calibrate the media pump (One of the HAT's five switched power outputs. In this build: 1 stirring, 2 media pump, 3 waste (product) pump, 4 CO₂ solenoid (relay), 5 heating. 3) too.
- If you see The media pump, not the waste pump: Calibrate the waste pump (PWM 2) too.
After running both pumps, what does the vial hold by weight?
Push the Media Out tube a little lower: it sets the level, and should be only just submerged at 14 ml.
Pull the Media Out tube up a little: it sets the level, and should be only just submerged at 14 ml.
Show the answer
Answer: 14 ml, the same as before the run
- If you see More than 14 ml: it has filled: Push the Media Out tube a little lower: it sets the level, and should be only just submerged at 14 ml.
- If you see Less than 14 ml: it has drained: Pull the Media Out tube up a little: it sets the level, and should be only just submerged at 14 ml.
Step checked